Double left ventricular pacing following accidental malpositioning of the right ventricular electrode during implantation of a cardiac resynchronization therapy device

在植入心脏再同步治疗装置过程中,由于右心室电极意外错位,导致双左心室起搏。

阅读:1

Abstract

Accidental malpositioning of a right ventricular (RV) electrode has not been previously reported in the context of cardiac resynchronization therapy (CRT).The case of a 75-year old male patient with dilative cardiomyopathy, left ventricular (LV) ejection fraction 23%, New York Heart Association functional heart failure status stage III, left bundle branch block (LBBB) with QRS width of 136 ms, and misplacement of the RV lead to the LV apex during implantation of a CRT defibrillator is described.Following unremarkable implantation, routine interrogation of the CRT device on the first day after the implantation revealed uneventful technical findings. The 12-lead surface electrocardiogram (ECG) showed biventricular stimulation featuring a narrow QRS complex with incomplete right bundle branch block (RBBB) and R>S in V1. The biplane postoperative chest X-ray was graded normal. On routine follow-up one month later, a transthoracic echocardiogram revealed an increased ejection fraction of 51% but the RV lead was placed in the LV apex. An additional transesophageal echocardiogram exhibited an Eustachian valve guiding the lead via the patent foramen ovale through the mitral valve into the LV apex. Operative revision was scheduled and the active fixation lead was uneventful removed from the LV. A new electrode was inserted and placed in the RV apex.Accidental malplacement of the RV electrode to the LV may be difficult to diagnose in the context of CRT patients as a stimulated biventricular ECG with incomplete RBBB appearance is expected in this situation. Careful analysis of lateral radiographic views during the operation is important to ensure correct lead positioning. As timely revision is the preferred procedure, early routine transthoracic echocardiography may be considered for detection of malplacement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。