A dosimetric comparison of stereotactic body radiation therapy techniques for lung cancer: robotic versus conventional linac-based systems

肺癌立体定向放射治疗技术的剂量学比较:机器人辅助系统与传统直线加速器系统

阅读:1

Abstract

The aim of this study is to compare the dosimetric characteristics of robotic and conventional linac-based SBRT techniques for lung cancer, and to provide planning guidance for each modality. Eight patients who received linac-based SBRT were retrospectively included in this study. A dose of 60 Gy given in three fractions was prescribed to each target. The Synchrony Respiratory Tracking System and a 4D dose calculation methodology were used for CyberKnife and linac-based SBRT, respectively, to minimize respiratory impact on dose calculation. Identical image and contour sets were used for both modalities. While both modalities can provide satisfactory target dose coverage, the dose to GTV was more heterogeneous for CyberKnife than for linac planning/delivery in all cases. The dose to 1000 cc lung was well below institutional constraints for both modalities. In the high dose region, the lung dose depended on tumor size, and was similar between both modalities. In the low dose region, however, the quality of CyberKnife plans was dependent on tumor location. With anteriorly-located tumors, the CyberKnife may deliver less dose to normal lung than linac techniques. Conversely, for posteriorly-located tumors, CyberKnife delivery may result in higher doses to normal lung. In all cases studied, more monitor units were required for CyberKnife delivery for given prescription. Both conventional linacs and CyberKnife provide acceptable target dose coverage while sparing normal tissues. The results of this study provide a general guideline for patient and treatment modality selection based on dosimetric, tumor and normal tissue sparing considerations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。