HbA1c and Glucose Management Indicator Discordance Associated with Obesity and Type 2 Diabetes in Intermittent Scanning Glucose Monitoring System

间歇扫描血糖监测系统中,HbA1c 和血糖管理指标不一致与肥胖和 2 型糖尿病相关

阅读:1

Abstract

Glucose management indicator (GMI) is frequently used as a substitute for HbA1c, especially when using telemedicine. Discordances between GMI and HbA1c were previously mostly reported in populations with type 1 diabetes (T1DM) using real-time CGM. Our aim was to investigate the accordance between GMI and HbA1c in patients with diabetes using intermittent scanning CGM (isCGM). In this retrospective cross-sectional study, patients with diabetes who used isCGM >70% of the time of the investigated time periods were included. GMI of four different time spans (between 14 and 30 days), covering a period of 3 months, reflected by the HbA1c, were investigated. The influence of clinical- and isCGM-derived parameters on the discordance was assessed. We included 278 patients (55% T1DM; 33% type 2 diabetes (T2DM)) with a mean HbA1c of 7.63%. The mean GMI of the four time periods was between 7.19% and 7.25%. On average, the absolute deviation between the four calculated GMIs and HbA1c ranged from 0.6% to 0.65%. The discordance was greater with increased BMI, a diagnosis of T2DM, and a greater difference between the most recent GMI and GMI assessed 8 to 10 weeks prior to HbA1c assessment. Our data shows that, especially in patients with increased BMI and T2DM, this difference is more pronounced and should therefore be considered when making therapeutic decisions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。