Abstract
Objective The optimal plate for fixation of tarsometatarsal joint injuries is controversial. The objective of this study was to compare the biomechanical characteristics between a locking plate and 1/4 tubular plate for first tarsometatarsal joint fracture-dislocation. Method Finite element analysis was used after establishment of a first tarsometatarsal joint fracture-dislocation model. Two implant simulations using a locking plate and five-hole 1/4 tubular plate were designed to simulate fixation of the fracture-dislocation. The displacement of the first tarsometatarsal articular surface and the stress distribution in the implants were calculated. Results A 700-N load was applied to both models. The minimum displacement of the articular surface in the locking plate and 1/4 tubular plate model was 0.6471 mm and 0.3833 mm, respectively. The maximum principal stress in the locking plate and 1/4 tubular plate was 1.212 × 10(3 )MPa and 1.107 × 10(3 )MPa, respectively. Conclusion Use of a 1/4 tubular plate is recommended for fixation of first tarsometatarsal joint fracture-dislocation after consideration of other factors such as economical issues.