Case report: False-positive B-type natriuretic peptide results with the cyclic amplified fluorescence immunoassay

病例报告:循环扩增荧光免疫测定法检测B型钠尿肽出现假阳性结果

阅读:1

Abstract

B-type natriuretic peptide, which was initially isolated from porcine brain tissue, has been used in the clinic as an important indicator for the diagnosis, risk stratification, and prognosis of heart failure. In recent years, many commercial immunoassays to assess B-type natriuretic peptide have been reported to undergo interference due to various factors. At present, the commonly used assays to assess B-type natriuretic peptide are enzyme-linked immunosorbent assay and direct chemiluminescence, which are based on a "sandwich structure" formed by dual antibodies; both assays have been reported to undergo interference. We report the case of a patient whose B-type natriuretic peptide results, which were detected using a cyclic amplified fluorescence immunoassay, were false-positive. The patient had a low-risk pulmonary embolism without right heart failure; however, her B-type natriuretic peptide results were surprisingly high. Further study confirmed that the abnormal B-type natriuretic peptide result was caused by the interference of the human anti-mouse antibodies. Therefore, when encountering B-type natriuretic peptide results that are not consistent with the clinical situation, the possibility of interference in the assay should be considered.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。