Intraocular pressure correlation to progressive retinal nerve fiber layer loss in primary open angle glaucoma as measured by standard and modified goldmann applanation tonometers

采用标准和改良的戈尔德曼压平式眼压计测量的原发性开角型青光眼眼压与进行性视网膜神经纤维层丢失的相关性

阅读:1

Abstract

PURPOSE: Characterize the relationship between intraocular pressure (IOP) as measured by standard and modified Goldmann prisms and the progressive loss of retinal nerve fiber layer (RNFL) in a cohort of glaucoma patients. DESIGN: Retrospective cross-sectional cohort data analysis. PARTICIPANTS: The study included all patients from a database of 1927 eyes, 966 patients with same visit sequential standard and modified Goldmann IOP measurements. From the database, 148 eyes, 75 patients met the inclusion criteria of a diagnosis of primary open angle glaucoma (POAG) with at least 5 sequential quality optical coherence tomographer (OCT) measurements. METHODS: Sequential OCT images were obtained with the spectral domain Zeiss OCT5000. Participants were all diagnosed with POAG by untreated IOP ≥ 22, disk changes, and visual field (HVF) loss consistent with glaucomatous optic neuropathy (GON). Included were 575 Goldmann IOP measurements with standard and modified prisms affixed to the Goldmann applanation tonometer (GAT) armature. A modified prism includes a corneal conforming applanation surface minimizing the cornea's contribution to the IOP measurement. The study included a total of 940 OCT visits with an average of 6.3 visits per eye over an average of 4.9 years. Retinal nerve fiber layer (RNFL) loss rate was calculated by serial linear fit of average RNFL thickness measurements. Demographics as well as central corneal thickness (CCT) and corneal hysteresis (CH) data were also collected. OUTCOME MEASURES: Pearson correlation coefficients and random coefficient models were used to evaluate the relationship between mean standard and modified IOP measurements and RNFL thickness measurements over time in POAG subjects. Secondary outcomes of CCT and CH correlation to RNFL were similarly analyzed. RESULTS: For all 148 POAG eyes, the overall rate of RNFL loss for an average standard GAT IOP of 17.9 mmHg was 1.08 µm per year (p = 0.002). Each 1-mmHg increase in standard GAT IOP was associated with an additional RNFL loss of 0.047 µm per year (r = 0.153, p = 0.06). Each 1-mmHg increase in modified GAT IOP was associated with an additional RNFL loss of 0.084 µm per year (r = 0.289, p = 0.0005). A modified prism IOP measurement ≥ 22 mmHg indicates a 2.57 times greater probability of significant RNFL loss than a standard prism IOP measurement ≥ 22 mmHg, p < 0.0001. CONCLUSIONS: Higher levels of GAT IOP during follow-up were related to higher rates of progressive RNFL loss detected by optic nerve OCT in treated POAG. A modified GAT prism surface demonstrates a significantly increased sensitivity, reliability and differentiation to progressive RNFL loss when compared to a standard GAT prism measured IOP. PRéCIS: A modified applanation surface prism with a corneal conforming shape used on a Goldmann tonometer appears to be a more sensitive and reliable indicator of progressive glaucomatous optic neuropathy as measured by retinal nerve fiber layer changes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。