Video analysis of optic-haptic-interaction during hydrophobic acrylic intraocular lens implantation using preloaded injectors

使用预装注射器对疏水性丙烯酸人工晶状体植入过程中的光学-触觉交互作用进行视频分析

阅读:1

Abstract

OBJECTIVE: To compare the optic-haptic interaction of different hydrophobic acrylic IOLs after using six preloaded injectors. METHODS: We reviewed the video-recordings of procedures on a total of 388 eyes that underwent phacoemulsification and intraocular lens (IOL) implantation. For six preloaded injectors: multiSert (Hoya Surgical Optics) [System 1], TECNIS Simplicity (Johnson & Johnson Vision) [System 2], TECNIS iTec (Johnson & Johnson Vision) [System 3], AutonoMe (Alcon, Laboratories) [System 4], Bluesert (Carl Zeiss Meditec) [System 5], and Prosert (OphthalmoPro GmbH) [System 6], we noted in each case the time of IOL delivery and made a descriptive observation of IOL insertion and optic-haptic-interaction. RESULTS: We defined standard haptic behavior where the haptics emerged "folded" from the injector and quickly recovered their pre-implantation appearance. The incidence where the leading haptic emerged in a deformed way for System 1 was 20%, System 2: 19%, System 3: 14%, System 4: 56%, System 5: 24% and System 6: 5%. For trailing haptic deformed behavior, the incidence was 36%, 6%, 4%, 8%, 18% and 2%, respectively for Systems 1 to 6. Optic-haptic adhesion occurred in 2% of cases for System 1, 44% for System 2, 52% for System 3, 48% for System 4, and 11% for System 6 (P < 0.05). Adhesion was not found with System 5. CONCLUSIONS: We observed different deformed behavior for leading and trailing haptics in the six preloaded systems, some systems had as much as 52% optic-haptic adhesion.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。