Influence of Selective Laser Trabeculoplasty (SLT) on the iStent inject® outcomes

选择性激光小梁成形术 (SLT) 对 iStent inject® 疗效的影响

阅读:1

Abstract

BACKGROUND: To evaluate the influence of Selective Laser Trabeculoplasty (SLT) on iStent inject® outcomes in open-angle glaucoma (OAG). METHODS: In this retrospective comparative cohort outcome study, 66 patients who were treated with two iStent inject® devices were included. Patients were divided into two subgroups consisting of patients without SLT treatment prior to surgery and patients who had been treated previously with 360° SLT but without sufficient response. Outcome measures included intraocular pressure (IOP) and number of antiglaucoma medications after 6 weeks with three, six, 12, and 24 month follow-ups. RESULTS: Mean preoperative IOP decreased from 20.4 ± 5.3 mmHg to 14.8 ± 3.0 mmHg for patients without SLT treatment prior to surgery (p = 0.001) and from 19.2 ± 4.5 mmHg to 14.0 ± 1.6 mmHg for patients with insufficient response to 360° SLT treatment (p = 0.027) at 12 months after iStent inject® implantation. No significant difference was found between the two groups (p >  0.05). The number of antiglaucoma medications did not change in both groups (p >  0.05) and showed no significant difference between the two groups (p >  0.05). CONCLUSION: Prior SLT treatment seems to have no negative influence on the IOP lowering-effect of iStent inject® implantation in patients with OAG. It is therefore an appropriate incremental procedure with no exclusion criterion for an iStent inject® implantation.