Goldmann and error correcting tonometry prisms compared to intracameral pressure

Goldmann 和误差校正眼压计棱镜与前房内压的比较

阅读:1

Abstract

BACKGROUND: Compare Goldmann applanation tonometer (GAT) prism and correcting applanation tonometry surface (CATS) prism to intracameral intraocular pressure (IOP), in vivo and in vitro. METHODS: Pressure transducer intracameral IOP was measured on fifty-eight (58) eyes undergoing cataract surgery and the IOP was modulated manometrically to 10, 20, and 40 mmHg. Simultaneously, IOP was measured using a Perkins tonometer with a standard GAT prism and a CATS prism at each of the intracameral pressures. Statistical comparison was made between true intracameral pressures and the two prism measurements. Differences between the two prism measurements were correlated to central corneal thickness (CCT) and corneal resistance factor (CRF). Human cadaver eyes were used to assess measurement repeatability. RESULTS: The CATS tonometer prism measured closer to true intracameral IOP than the GAT prism by 1.7+/-2.7 mmHg across all pressures and corneal properties. The difference in CATS and GAT measurements was greater in thin CCT corneas (2.7+/-1.9 mmHg) and low resistance (CRF) corneas (2.8+/-2.1 mmHg). The difference in prisms was negligible at high CCT and CRF values. No difference was seen in measurement repeatability between the two prisms. CONCLUSION: A CATS prism in Goldmann tonometer armatures significantly improve the accuracy of IOP measurement compared to true intracameral pressure across a physiologic range of IOP values. The CATS prism is significantly more accurate compared to the GAT prism in thin and less rigid corneas. The in vivo intracameral study validates mathematical models and clinical findings in IOP measurement between the GAT and CATS prisms.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。