Comparison of CyTOF assays across sites: Results of a six-center pilot study

不同中心CyTOF检测方法的比较:一项六中心试点研究的结果

阅读:12
作者:Michael D Leipold ,Gerlinde Obermoser ,Craig Fenwick ,Katja Kleinstuber ,Narges Rashidi ,John P McNevin ,Allison N Nau ,Lisa E Wagar ,Virginie Rozot ,Mark M Davis ,Stephen DeRosa ,Giuseppe Pantaleo ,Thomas J Scriba ,Bruce D Walker ,Lars R Olsen ,Holden T Maecker

Abstract

For more than five years, high-dimensional mass cytometry has been employed to study immunology. However, these studies have typically been performed in one laboratory on one or few instruments. We present the results of a six-center study using healthy control human peripheral blood mononuclear cells (PBMCs) and commercially available reagents to test the intra-site and inter-site variation of mass cytometers and operators. We used prestained controls generated by the primary center as a reference to compare against samples stained at each individual center. Data were analyzed at the primary center, including investigating the effects of two normalization methods. All six sites performed similarly, with CVs for both Frequency of Parent and median signal intensity (MSI) values<30%. Increased background was seen when using the premixed antibody cocktail aliquots at each site, suggesting that cocktails are best made fresh. Both normalization methods tested performed adequately for normalizing MSI values between centers. Clustering algorithms revealed slight differences between the prestained and the sites-stained samples, due mostly to the increased background of a few antibodies. Therefore, we believe that multicenter mass cytometry assays are feasible. Keywords: Clustering; CyTOF; Mass cytometry; Multicenter; Normalization; Standardization.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。