Platinum stable isotope analysis of geological standard reference materials by double-spike MC-ICPMS

采用双同位素稀释法MC-ICPMS分析地质标准参考物质中的铂稳定同位素。

阅读:1

Abstract

We report a method for the chemical purification of Pt from geological materials by ion-exchange chromatography for subsequent Pt stable isotope analysis by multiple-collector inductively coupled plasma mass spectrometry (MC-ICPMS) using a (196)Pt-(198)Pt double-spike to correct for instrumental mass bias. Double-spiking of samples was carried out prior to digestion and chemical separation to correct for any mass-dependent fractionation that may occur due to incomplete recovery of Pt. Samples were digested using a NiS fire assay method, which pre-concentrates Pt into a metallic bead that is readily dissolved in acid in preparation for anion-exchange chemistry. Pt was recovered from anion-exchange resin in concentrated HNO(3) acid after elution of matrix elements, including the other platinum group elements (PGE), in dilute HCl and HNO(3) acids. The separation method has been calibrated using a precious metal standard solution doped with a range of synthetic matrices and results in Pt yields of ≥90% with purity of ≥95%. Using this chemical separation technique, we have separated Pt from 11 international geological standard reference materials comprising of PGE ores, mantle rocks, igneous rocks and one sample from the Cretaceous-Paleogene boundary layer. Pt concentrations in these samples range from ca. 5 ng g(-1) to 4 μg g(-1). This analytical method has been shown to have an external reproducibility on δ(198)Pt (permil difference in the (198)Pt/(194)Pt ratio from the IRMM-010 standard) of ±0.040 (2 sd) on Pt solution standards (Creech et al., 2013, J. Anal. At. Spectrom. 28, 853-865). The reproducibility in natural samples is evaluated by processing multiple replicates of four standard reference materials, and is conservatively taken to be ca. ±0.088 (2 sd). Pt stable isotope data for the full set of reference materials have a range of δ(198)Pt values with offsets of up to 0.4‰ from the IRMM-010 standard, which are readily resolved with this technique. These results demonstrate the potential of the Pt isotope system as a tracer in geochemical systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。