Does a novel 3D printed individualized guiding template based on cutaneous fiducial markers contribute to accurate percutaneous insertion of pelvic screws? A preliminary phantom and cadaver study

一种基于皮肤定位标记的新型3D打印个性化导向模板能否提高经皮骨盆螺钉置入的准确性?一项初步的体模和尸体研究

阅读:1

Abstract

BACKGROUND: Most 3D-printed guiding templates require dissection of soft tissues to match the corresponding surfaces of the guiding templates. This study sought to explore the accuracy and acceptability of the novel 3D printed individualized guiding templates based on cutaneous fiducial markers in minimally invasive screw placement for pelvic fractures. METHODS: The printed template was tested on five high-fidelity biomimetic phantom models of the bony pelvis and its surrounding soft tissues as well as on two fresh frozen cadavers. Four cutaneous fiducial markers were transfixed on each phantom model prior to performing CT scans to reconstruct their 3D models. Personalized templates for guiding screw insertion were designed based on the positions of the fiducial markers and virtually planned target screw channels after scanning, followed by 3D printing of the guide. Phase 1 consisted of five expert surgeons inserting one anterograde supra-pubic screw and one sacroiliac screw percutaneously into each phantom model using the 3D-printed guide. The deviation of screw positions between the pre-operative planned and post-operative actual ones was measured after registering their 3D modelling. A Likert scale questionnaire was completed by the expert surgeons to assess their satisfaction and acceptability with the guiding template. Phase 2 consisted of repeating the same procedures on the fresh frozen cadavers in order to demonstrate face, content and concurrent validity. RESULTS: In Phase 1, all ten screws were successfully implanted with the assistance of the guiding template. Postoperative CT scans confirmed that all screws were safely positioned within the bony pelvic channels without breaching the far cortex. The mean longitudinal deviation at the bony entry point and screw tip between the pre-operative planned and post-operative actual screw paths were 2.83 ± 0.60 mm and 3.12 ± 0.81 mm respectively, with a mean angular deviation of 1.25 ± 0.41°. Results from the Likert questionnaire indicated a high level of satisfaction for using the guiding template among surgeons. In Phase 2, results were similar to those in Phase 1. CONCLUSIONS: The 3D-printed guiding template based on cutaneous fiducial markers shows potential for assisting in the accurate insertion of percutaneous screws in the pelvis.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。