The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a (192)Ir HDR Brachytherapy Source

利用电离室测量(192)Ir高剂量率近距离放射治疗源在空气和固体模型中的空气比释动能率

阅读:1

Abstract

INTRODUCTION: This study aims to measure the air-kerma rate of 192-Ir-HDR-afterloading source with an ionization chamber in air and a solid cylindrical phantom separately and to compare the dose calibration by the American Association of Physicists in Medicine (AAPM) Task Group TG-43U1 formalism with the Abacus treatment planning system (TPS). MATERIALS AND METHODS: The air-kerma rate of (192)Ir source was measured by an ionization chamber in air and a solid cylindrical phantom separately. For the interesting point position P (8cm, 90°), the values of the dose were calculated with the TG-43U1 formula and compared with data from the Abacus TPS with single and multiple dwell positions, respectively. RESULTS: The air-kerma rate percentage deviations between the detector measurements in air and the source certificate were -1.28%, -0.91%, -0.71%, and 0.33% at the distances of 25cm, 50cm, 75cm, and 100cm, respectively. For the measurement in solid cylindrical phantom, the percent deviation from the air-kerma rate certificate was 1.85%. The percentage deviations of the dose calibration between Abacus TPS and TG-43U1 formalism at P (8cm, 90°) were -2.30%, 1.76%, and 2.10% with different distances (between the dwell positions) of 0cm, 0.5cm, and 1cm, respectively. CONCLUSION: The in-air technique was a new attempt for clinic routine measurement. Further studies are still necessary. As a treatment planning system, the Abacus TPS should apply the AAPM TG-43U1 formulism for the development required in the future.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。