Proprieties of adhesive surface arrays to thyroid cartilage for recurrent laryngeal nerve monitoring

用于喉返神经监测的粘附表面阵列在甲状软骨上的应用特性

阅读:1

Abstract

BACKGROUND: To investigate the optimal placement and size of adhesive thyroid cartilage electrodes (ATCEs) placed on the thyroid cartilage in porcine models for evaluating recurrent laryngeal nerve function compared with the other 3 recording electrode types. METHODS: Four Meishan piglets were used for this study. The electromyogram (EMG) signal stability was detected by intraoperative nerve monitoring (IONM). Best location and size of ATCEs were detected, and the EMG signals from each were compared. Latency data were tested by stimulating the different nerve points, and 6 manipulations of the trachea and endotracheal tube (ET) were applied to test the stability of EMG signals. ET electrodes, needle electrodes, and transcutaneous recording electrodes were simultaneously tested with comparison to ATCEs. RESULTS: The optimal placement locations and sizes of ATCEs were determined. The amplitudes and latencies recorded from the ATCEs were consistent with those of the ET electrodes. More anti-interference was observed with the ATCEs than with the ET electrodes in surgical manipulations. ATCEs could be used during intermittent and continuous monitoring in similar fashion to ET electrodes and needle electrodes. CONCLUSIONS: ATCEs had consistent monitoring function with ET electrodes and needle electrodes. Feasibility, EMG stability, and optimal location and size of ATCEs for IONM were resolved. The EMG profiles from the ATCEs were more stable during surgical manipulations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。