Direct Coupling of Bio-SPME to Liquid Electron Ionization-MS/MS via a Modified Microfluidic Open Interface

通过改进的微流体开放接口将 Bio-SPME 与液体电子电离-MS/MS 直接耦合

阅读:4
作者:Priscilla Rocío-Bautista, Giorgio Famiglini, Veronica Termopoli, Pierangela Palma, Emir Nazdrajić, Janusz Pawliszyn, Achille Cappiello

Abstract

We present a modified microfluidic open interface (MOI) for the direct coupling of Bio-SPME to a liquid electron ionization-tandem mass spectrometry (LEI-MS/MS) system as a sensitive technique that can directly analyze biological samples without the need for sample cleanup or chromatographic separations as well as without measurable matrix effects (ME). We selected fentanyl as test compound. The method uses a C18 Bio-SPME fiber by direct immersion (DI) in urine and plasma and the subsequent quick desorption (1 min) in a flow-isolated volume (2.5 μL) filled with an internal standard-acetonitrile solution. The sample is then transferred to an EI source of a triple-quadrupole mass spectrometer via a LEI interface at a nanoscale flow rate. The desorption and analysis procedure requires less than 10 min. Up to 150 samples can be analyzed without observing a performance decline, with fentanyl quantitation at microgram-per-liter levels. The method workflow is extremely dependable, relatively fast, sustainable, and leads to reproducible results that enable the high-throughput screening of various biological samples.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。