Task-driven source-detector trajectories in cone-beam computed tomography: II. Application to neuroradiology

锥束计算机断层扫描中任务驱动的源-探测器轨迹:II. 在神经放射学中的应用

阅读:1

Abstract

We apply the methodology detailed in "Task-driven source-detector trajectories in cone-beam computed tomography: I. Theory and methods" by Stayman et al. for task-driven optimization of source-detector orbits in cone-beam computed tomography (CBCT) to scenarios emulating imaging tasks in interventional neuroradiology. The task-driven imaging framework is used to optimize the CBCT source-detector trajectory by maximizing the detectability index, d' . The approach was applied to simulated cases of endovascular embolization of an aneurysm and arteriovenous malformation and was translated to real data first using a CBCT test bench followed by implementation on an interventional robotic C-arm. Task-driven trajectories were found to generally favor higher fidelity (i.e., less noisy) views, with an average increase in d' ranging from 7% to 28%. Visually, this resulted in improved conspicuity of particular stimuli by reducing the noise and altering the noise correlation to a form distinct from the spatial frequencies associated with the imaging task. The improvements in detectability and the demonstration of the task-driven workflow using a real interventional imaging system show the potential of the task-driven imaging framework to improve imaging performance on motorized, multiaxis C-arms in neuroradiology.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。