Development of a noncontact diffuse optical spectroscopy probe for measuring tissue optical properties

用于测量组织光学特性的非接触式漫射光谱探头的开发

阅读:9
作者:Sheldon F Bish, Narasimhan Rajaram, Brandon Nichols, James W Tunnell

Abstract

Optical reflectance probes are often used as tools to obtain optical spectra from superficial tissues and subsequently determine optical and physiological properties associated with early stage cancer. These probes, when placed directly on the tissue, are known to cause significant pressure-dependent changes in local optical properties. To address this, we fit the probe with an optical device that images the illumination and collection fibers onto the tissue surface, eliminating the influence of contact probe pressure on the sampling area. The noncontact probe addition addresses new optical conditions that may affect its performance such as tissue surface contour, and specular reflections by implementing an autofocusing mechanism and cross polarization. Extracted optical properties of tissue simulating phantoms yield errors of 3.46% in reduced scattering and 8.62% in absorbance. Autofocusing has extended the depth of field from 4 mm to throughout the 12 mm range of autofocus travel, while cross polarization has removed the incidence angle dependent specular reflection component from the collected signal.