Development and validation of a multiplexed LC-MS/MS ketone body assay for clinical diagnostics

用于临床诊断的多重 LC-MS/MS 酮体检测方法的开发和验证

阅读:10

Conclusions

The developed method is highly sensitive and robust. It has been validated for use with human serum and plasma, overcoming stability concerns and providing a reliable and efficient quantitative estimation of ketone bodies.

Methods

A novel 6.5-minute liquid chromatography-mass spectrometry-based assay was developed, enabling the precise measurement of alpha-, beta- and gamma-hydroxybutyrate, beta-hydroxyisobutyrate, and acetoacetate. This method was fully validated for human serum and plasma samples by investigating extraction efficiency, matrix effects, accuracy, recovery, intra- and inter-precision, linearity, lower limit of quantitation (LLOQ), carryover, specificity, stability, and more. From 107 normal samples, reference ranges were established for all analytes and the beta-hydroxybutyrate/acetoacetate ratio.

Results

All five analytes were adequately separated chromatographically. An extraction efficiency between 80 and 120 % was observed for all KBs. Accuracy was evaluated through spike and recovery using 10 random patient samples, with an average recovery of 85-115 % for all KBs and a coefficient of variation of ≤ 3 %. Coefficients of variation for intra- and inter-day imprecision were < 5 %, and the total imprecision was < 10 %. No significant interferences were observed. Specimens remained stable for up to 6 h on ice or 2 h at room temperature. Conclusions: The developed method is highly sensitive and robust. It has been validated for use with human serum and plasma, overcoming stability concerns and providing a reliable and efficient quantitative estimation of ketone bodies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。