Method validation of multi-element panel in whole blood by inductively coupled plasma mass spectrometry (ICP-MS)

电感耦合等离子体质谱法(ICP-MS)对全血中多元素组进行方法验证

阅读:12
作者:Amol O Bajaj, Rebecca Parker, Candice Farnsworth, Christian Law, Kamisha L Johnson-Davis

Background

Analytical

Conclusions

The multi-element panel was developed to analyze ten elements in whole blood to unify the sample preparation and increase batch run efficiency. The improved analytical method utilized matrix-matched calibrators for accurate quantification to meet regulatory requirements. The assay was validated according to guidelines for CLIA-certified clinical laboratories and was suitable for clinical testing to assess nutritional status and toxic exposure.

Methods

A method was developed and validated to quantify ten elements in whole blood by ICP-MS. Fifty microliters of sample were extracted with 950 μL of diluent containing 1 % ammonium hydroxide, 0.1 % Triton X-100, 1.75 % EDTA along with spiked internal standards. Four calibrators were used for each element and prepared in goat blood to match the patient specimen matrix. Samples were analyzed with an Agilent 7700 ICP-MS with a Cetac MVX 7100 μL Workstation autosampler.

Results

The assay was linear for all elements with inter- and intra-assay imprecision less than or equal to 11% CV at the low end of the analytical measurement range (AMR) and less than or equal to 4% CV at the upper end of the AMR for all elements. Accuracy was checked with a minimum of 40 repeat patient samples, proficiency testing samples, and matrix-matched spikes. The linear slopes for the ten elements ranged from 0.94 to 1.03 with intercepts below the AMR and R2 ranging from 0.97 to 1.00. Conclusions: The multi-element panel was developed to analyze ten elements in whole blood to unify the sample preparation and increase batch run efficiency. The improved analytical method utilized matrix-matched calibrators for accurate quantification to meet regulatory requirements. The assay was validated according to guidelines for CLIA-certified clinical laboratories and was suitable for clinical testing to assess nutritional status and toxic exposure.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。