Migration and section of esophageal thermometer in laparoscopic sleeve gastrectomy: Clinical case of a preventable complication

腹腔镜袖状胃切除术中食管温度计移位和断裂:一例可预防并发症的临床病例

阅读:2

Abstract

INTRODUCTION AND IMPORTANCE: Laparoscopic sleeve gastrectomy is a prevalent bariatric surgery for managing morbid obesity. Despite its efficacy, complications can arise, particularly from intraoperative devices such as esophageal thermometers and orogastric tubes. These devices, if misplaced or inadequately monitored, can migrate and become entrapped or sectioned during surgery, leading to significant morbidity. CASE PRESENTATION: A 49-year-old female with morbid obesity underwent LSG. During surgery, an esophageal thermometer migrated into the stomach and was inadvertently sectioned during the stapling process. This required additional surgical intervention to remove the severed segments and repair the damage. Postoperative recovery was uneventful after corrective measures were taken. CLINICAL DISCUSSION: This case underscores the critical importance of ensuring the proper placement and continuous monitoring of intraoperative devices, such as esophageal thermometers, to prevent similar preventable complications in future surgical procedures. CONCLUSION: The reviewed cases demonstrate that complications from intraoperative device migration and entrapment during bariatric surgery, while rare, are significant and preventable. Adherence to strict protocols, continuous device monitoring, and enhanced team communication are essential to improve patient safety and surgical outcomes. Implementing these measures can prevent avoidable complications and enhance the efficacy of bariatric surgeries. EVIDENCE BASED MEDICINE RANKING: Level IV.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。