Evaluation of Milled Titanium versus Laser Sintered Co-Cr Abutments on the Marginal Misfit in Internal Implant-Abutment Connection

评估铣削钛基台与激光烧结钴铬合金基台在内连接种植体-基台连接边缘密合度方面的差异

阅读:1

Abstract

The precision of fit at the implant-abutment connection is an important criterion for the clinical success of restorations and implants. Several factors are involved among which are the abutment materials and manufacturing techniques. The aim of this study was to investigate the effect of two materials and methods of manufacturing implant abutments, milled titanium versus laser sintered Co-Cr, on the marginal misfit at the implant-abutment interface. Scanning electron microscopes (SEM) were used to geometrically measure the marginal vertical discrepancy of a total of 80 specimens, classified into eight categories, according to the implant system and abutment. The data were statistically analyzed by Student's paired t test, one-way and two-way ANOVA with the Bonferroni-Holm correction at the significance level of p = 0.05. Milled titanium abutments demonstrated the lowest misfit values in the implant systems analyzed. The marginal fit of all the groups was within the clinically acceptable range for implant prostheses.