Evaluation of BACTEC system for urine culture screening

对用于尿培养筛查的BACTEC系统进行评估

阅读:1

Abstract

We compared urine culturing performed by the calibrated loop method with a screening system (BACTEC). A total of 852 urine specimens were examined by both the conventional loop method and the BACTEC system. With the loop method, 193 (22.6%) urine samples were positive, whereas with the BACTEC system, 185 (21.7%) were positive (sensitivity, 96.01%). At a breakpoint of 10(4) CFU/ml, eight false-negatives were detected (sensitivity, 87.09%), and at a breakpoint of 10(5) CFU/ml, four false-negatives were observed (sensitivity, 97.6%). The specificity of the BACTEC system was 100%. We propose the BACTEC method as an effective alternative to the other growth-dependent screening tests for bacteriuria.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。