When battery exhaustion lets the lame walk: a case report on the importance of long-term stimulator monitoring in deep brain stimulation

当电池耗尽让跛子也能行走:关于深部脑刺激中长期刺激器监测重要性的病例报告

阅读:1

Abstract

BACKGROUND: Deep brain stimulation is increasingly used in the treatment of advanced Parkinson's disease. While its short-term effectiveness is well documented, there are only few reports on long-term outcomes, and the need to repeatedly reprogram the stimulator is seldom reported. CASE PRESENTATION: We present a 74-year-old man with gait impairment, which had been mistaken for worsening of the disease and only remitted when the stimulator battery was exhausted indicating that the stimulator itself had been the cause. CONCLUSION: This case highlights the need to repeatedly monitor not only battery capacity, but also stimulator-related side-effects for an extended period after implantation and, if necessary, to refer to centres capable of systematically reprogramming the device.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。