Quantification of emicizumab by mass spectrometry in plasma of people with hemophilia A: A method validation study

利用质谱法定量分析A型血友病患者血浆中艾美赛珠单抗的含量:方法验证研究

阅读:1

Abstract

BACKGROUND: Emicizumab is a new treatment option for people with hemophilia A. Emicizumab was approved with a body-weight-based dosage regimen, without laboratory monitoring requirements. Guidelines, however, recommend measuring emicizumab concentrations when the presence of antidrug antibodies is suspected. Furthermore, drug monitoring can be useful in clinical decision making, in adherence checking, and for research purposes. Therefore, we developed a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for quantifying emicizumab. We performed a validation study on this LC-MS/MS method quantifying emicizumab in the plasma of people with hemophilia A. METHODS: Sample preparation for LC-MS/MS analysis included ammonium sulfate protein precipitation and trypsin digestion. A signature peptide of emicizumab and a matching stable isotope-labeled internal standard were used to quantify emicizumab by LC-MS/MS analysis. Validation was performed in accordance with the "Guideline on Bioanalytical Method Validation" of the European Medicines Agency (EMA). The LC-MS/MS method was cross validated against a modified and calibrated (r (2) Diagnostics) one-stage clotting assay (OSA). CONCLUSIONS: The LC-MS/MS method demonstrated linearity over a wide range of emicizumab concentrations, far exceeding the concentrations observed in people with hemophilia A. Precision and accuracy were excellent, and all other validation parameters were also within the acceptance EMA criteria. Cross validation showed that the LC-MS/MS method and the OSA-based method can be used interchangeably for drug monitoring of emicizumab without the application of a correction factor.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。