Biomechanical compression tests with a new implant for thoracolumbar vertebral body replacement

使用新型胸腰椎椎体置换植入物进行生物力学压缩测试

阅读:1

Abstract

The authors present an investigation into the biomechanical functioning of a new titanium implant for vertebral body replacement (Synex). Possible indications are fractures and/or dislocations with damage of the anterior column, posttraumatic kyphosis and tumors of the thoracolumbar spine. The construction must be supplemented by a stabilizing posterior or anterior implant. For best fit and contact with adjacent end-plates, Synex is distractable in situ. We performed comparative compression tests with Synex and MOSS ("Harms mesh cage") on human cadaveric specimens of intact vertebrae (L1). The aim of the study was to measure the compressive strength of the vertebral body end-plate in uniaxial loading via both implants to exclude collapse of Synex in vivo. Twelve human cadaveric specimens of intact vertebrae (L1) were divided into two identical groups (matched pairs) according to bone mineral density (BMD), determined using dual-energy quantitative computed tomography (DE-QCT). The specimens were loaded with an axial compression force at a constant speed of 5 mm/min to failure, and the displacement was recorded with a continuous load-displacement curve. The mean ultimate compression force (Fmax) showed a tendency towards a higher reading for Synex: 3396 N versus 2719 N (non-significant). The displacement until Fmax was 2.9 mm in the Synex group, which was half as far as in the MOSS group (5.8 mm). The difference was significant (P < 0.001). The compression force was twice as high, and significantly (P < 0.05) higher with Synex at displacements of 1 mm, 1.5 mm and 2 mm. A significant (P < 0.001) correlation (R = 0.89) between Fmax and BMD was found. Synex was found to be at least comparable to MOSS concerning the compressive performance at the vertebral end-plate. A possible consequence of the significantly higher mean compression forces between 1 and 2 mm displacement might be decreased collapse of the implant into the vertebral body in vivo.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。