A combination of parabolic and grid slope interpolation for 2D tissue displacement estimations

二维组织位移估计的抛物线插值与网格斜率插值相结合

阅读:1

Abstract

Parabolic sub-sample interpolation for 2D block-matching motion estimation is computationally efficient. However, it is well known that the parabolic interpolation gives a biased motion estimate for displacements greater than |y.2| samples (y = 0, 1, …). Grid slope sub-sample interpolation is less biased, but it shows large variability for displacements close to y.0. We therefore propose to combine these sub-sample methods into one method (GS15PI) using a threshold to determine when to use which method. The proposed method was evaluated on simulated, phantom, and in vivo ultrasound cine loops and was compared to three sub-sample interpolation methods. On average, GS15PI reduced the absolute sub-sample estimation errors in the simulated and phantom cine loops by 14, 8, and 24% compared to sub-sample interpolation of the image, parabolic sub-sample interpolation, and grid slope sub-sample interpolation, respectively. The limited in vivo evaluation of estimations of the longitudinal movement of the common carotid artery using parabolic and grid slope sub-sample interpolation and GS15PI resulted in coefficient of variation (CV) values of 6.9, 7.5, and 6.8%, respectively. The proposed method is computationally efficient and has low bias and variance. The method is another step toward a fast and reliable method for clinical investigations of longitudinal movement of the arterial wall.