Usefulness of a finger-mounted tissue oximeter with near-infrared spectroscopy for evaluating the intestinal oxygenation and viability in rats

利用近红外光谱技术对大鼠指套式组织氧饱和度仪评估肠道氧合和活力的实用性

阅读:1

Abstract

PURPOSE: To investigate the utility of the device for evaluating intestinal oxygenation and viability using an animal model. METHODS: Sprague-Dawley rats underwent laparotomy under general anesthesia, and the blood vessels in the terminal ileum were clamped to create ischemia. We measured the regional tissue oxygenation saturation (rSO(2)) using an oximeter after 1, 3, and 6 h of vessel clamping. Ischemic tissue damage was assessed using a histological score. The intestine was reperfused after each clamping period, and intestinal rSO(2) and survival rate were evaluated. RESULTS: When reperfusion was performed at 1 and 3 h after ischemia, rSO(2) increased after 10 min, and it improved to the same level as for normal intestine after 1 h; all rats survived for 1 week. In contrast, after 6 h of ischemia, rSO(2) did not increase after reperfusion, and all animals died within 2 days. The histological scores increased after 1 h of reperfusion, with longer clamping periods. CONCLUSION: A finger-mounted tissue oximeter could evaluate intestinal ischemia and the viability, which is thus considered to be a promising result for future clinical application.