Surveillance of Arrhythmia in Patients After Myocardial Infarction Using Wearable Electrocardiogram Patch Devices: Prospective Cohort Study

利用可穿戴式心电图贴片设备监测心肌梗死后患者的心律失常:前瞻性队列研究

阅读:1

Abstract

BACKGROUND: Acute myocardial infarction may be associated with new-onset arrhythmias. Patients with myocardial infarction may manifest serious arrhythmias such as ventricular tachyarrhythmias or atrial fibrillation. Frequent, prolonged electrocardiogram (ECG) monitoring can prevent devastating outcomes caused by these arrhythmias. OBJECTIVE: We aimed to investigate the incidence of arrhythmias in patients following myocardial infarction using a patch-type device-AT-Patch (ATP-C120; ATsens). METHODS: This study is a nonrandomized, single-center, prospective cohort study. We evaluated 71 patients who had had a myocardial infarction and had been admitted to our hospital. The ATP-C120 device was attached to the patient for 11 days and analyzed by 2 cardiologists for new-onset arrhythmic events. RESULTS: One participant was concordantly diagnosed with atrial fibrillation. The cardiologists diagnosed atrial premature beats in 65 (92%) and 60 (85%) of 71 participants, and ventricular premature beats in 38 (54%) and 44 (62%) participants, respectively. Interestingly, 40 (56%) patients showed less than 2 minutes of sustained paroxysmal atrial tachycardia confirmed by both cardiologists. Among participants with atrial tachycardia, the use of β-blockers was significantly lower compared with patients without tachycardia (70% vs 90%, P=.04). However, different dosages of β-blockers did not make a significant difference. CONCLUSIONS: Wearable ECG monitoring patch devices are easy to apply and can correlate symptoms and ECG rhythm disturbances in patients following myocardial infarction. Further study is necessary regarding clinical implications and appropriate therapies for arrhythmias detected early after myocardial infarction to prevent adverse outcomes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。