Effect of Repeated Moist Heat Sterilization on Titanium Implant-Abutment Interface-An In Vitro Study

反复湿热灭菌对钛种植体-基台界面影响的体外研究

阅读:1

Abstract

OBJECTIVES:  Sterilization eliminates microbial viability by decreasing the biological load, but likewise have the ability to deteriorate the mechanical properties of an implant material. This study intended to evaluate the effect of repeated moist heat sterilization on implant-abutment interface using two different implant systems. MATERIALS AND METHODS:  Forty screw-retained titanium implant-abutment combinations (fixture 3.5 ×10 mm, abutment 2 mm diameter), twenty each from Genesis (Aktiv Implant Systems, United States) and Bredent (SKY, Germany), were divided into four different groups (n = 10) and placed in a computer-aided diagnostic model. The abutments from each group were exposed to first and second autoclave cycle (121°C for 30 minutes), connected back to the fixture and analyzed under scanning electron microscope for marginal gap and surface roughness. RESULTS:  Genesis group showed higher marginal gaps on both sides (buccal/mesial [2.8 ± 0.47]; lingual/distal [2.8 ± 0.33]), while Bredent implant-abutment system (IAS) did not show any changes in marginal gaps after autoclaving. Differences within and between the group were found to be statistically significant. Surface roughness for Genesis (243.7 ± 70.30) and Bredent groups (528.9 ± 213.19) was highest at second autoclave, with Bredent implant-abutment showing higher values for surface roughness than Genesis IAS. CONCLUSION:  Marginal vertical gap increased with autoclaving for Genesis IAS, while Bredent implant abutments were more stable. Surface roughness increases with autoclaving for both Genesis and Bredent group of IAS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。