Development of a new UHV/XHV pressure standard (Cold Atom Vacuum Standard)

开发新的超高真空/超高真空压力标准(冷原子真空标准)

阅读:1

Abstract

The National Institute of Standards and Technology has recently begun a program to develop a primary pressure standard that is based on ultra-cold atoms, covering a pressure range of 1 × 10(-6) Pa to 1 × 10(-10) Pa and possibly lower. These pressures correspond to the entire ultra-high vacuum (UHV) range and extend into the extreme-high vacuum (XHV). This cold-atom vacuum standard (CAVS) is both a primary standard and absolute sensor of vacuum. The CAVS is based on the loss of cold, sensor atoms (such as the alkali-metal lithium) from a magnetic trap due to collisions with the background gas (primarily H(2)) in the vacuum. The pressure is determined from a thermally-averaged collision cross section, which is a fundamental atomic property, and the measured loss rate. The CAVS is primary because it will use collision cross sections determined from ab initio calculations for the Li + H(2) system. Primary traceability is transferred to other systems of interest using sensitivity coefficients.