Achievement of 0.005 % combined transfer uncertainties in the NIST detector calibration facility

NIST探测器校准设施实现了0.005%的综合传递不确定度

阅读:1

Abstract

Improvements in a lamp-monochromator-based facility at the National Institute of Standards and Technology (NIST), the Visible near-infrared Spectral Comparator Facility (VisSCF) which is used to calibrate optical detectors for spectral radiant power responsivity from 300 nm to 1100 nm, are described. These changes include extending the VisSCF operational range down to 300 nm from 350 nm, thereby fully covering the ultraviolet-A (UVA) spectral region and partially covering the UVB range. These improvements have lowered the magnitudes of most of the components in the uncertainty budget and have led to combined 0.005 % transfer (k=1) uncertainties in the spectral power responsivity calibrations over most of the spectral range. Redevelopment of the uncertainty budget results in total expanded uncertainties of spectral responsivities of less than 0.1 % (k=2) over the spectral range from 380 nm to 980 nm, with the greatest uncertainty term coming from the calibrations of the transfer standards.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。