Final report on the key comparison, CCM.P-K15 in the pressure range from 1.0 × 10(-4) Pa to 1.0 Pa

关键对比的最终报告,CCM.P-K15,压力范围为 1.0 × 10(-4) Pa 至 1.0 Pa

阅读:1

Abstract

The comparison CCM.P-K15 is a key comparison in pressure involving six laboratories in three regional metrological organizations (RMO). The measurand of the comparison is the accommodation coefficient of two spinning rotating gauge characterized in nitrogen from 0.1 mPa up to 1.0 Pa. The two transfer standards were circulated from November 2009 until March 2011. The circulation consisted of three loops, one for each RMO, and a new calibration by the pilot between each loop. The stability of one of the transfer standards was poor and was worse than expected based on the previous history of the transfer standard while the other transfer standard demonstrated good stability while circulated in Europe and America and a fair stability while circulated in Asia. All the participants demonstrated equivalence to the definition of pressure in their respective primary standards.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。