Study of Nickel Levels in Patients with Atrial Septal Defect Undergoing Amplatzer Device Closure

对接受Amplatzer装置封堵术的房间隔缺损患者镍水平的研究

阅读:1

Abstract

Transcatheter closure of atrial septal defects (ASDs) using nickel titanium Naval Ordnance Laboratory (Nitinol)-containing devices (Amplatzer, AGA Medical, Golden Valley, Minnetosa) is being widely practiced. Controversies still exist regarding the release of nickel from these devices and the allergy and other ill effects of nickel and also the duration of antiplatelet therapy to aid endothelization of the device. OBJECTIVE: To study the nickel levels in patients who underwent ASD closure with platinum-coated Nitinol-containing Amplatzer septal occluder. METHODOLOGY: A prospective study was conducted on 25 patients, between 6 years and 40 years of age who underwent ASD closure with Amplatzer septal occlude sizes ranging from 12 to 36 mm from January 2009 to January 2010. Blood nickel levels were estimated using atomic absorption photometry before and 24 h after the procedure and later at 1 month, 3 months, and 6 months postprocedure. A value of <2 mcg/dl was considered to be normal. Statistical analysis was performed by the use of the Wilcoxon test. RESULTS: The blood nickel levels at mean baseline, 24 h, 1 month, 3 months, and 6 months postprocedure were 1.05, 1.39, 0.98, 0.79, and 0.74 (mcg/dl), respectively. CONCLUSIONS: Percutaneous ASD closure using Nitinol devices can be carried out safely without any significant ill effects related to nickel release.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。