First-in-Human Study for Evaluating the Accuracy of Smart Ring Based Cuffless Blood Pressure Measurement

首次人体研究评估基于智能戒指的无袖带血压测量的准确性

阅读:1

Abstract

BACKGROUND: Recently, a ring-type cuffless blood pressure (BP) measuring device has been developed. This study was a prospective, single arm, first-in-human pivotal trial to evaluate accuracy of BP measurement by the new device. METHODS: The ring-type smart wearable monitoring device measures photoplethysmography signals from the proximal phalanx and transmits the data wirelessly to a connected smartphone. For the BP comparison, a cuff was worn on the arm to check the reference BP by auscultatory method, while the test device was worn on the finger of the opposite arm to measure BP simultaneously. Measurements were repeated for up to three sets each on the left and right arms. The primary outcome measure was mean difference and standard deviation of BP differences between the test device and the reference readings. RESULTS: We obtained 526 sets of systolic BP (SBP) and 513 sets of diastolic BP (DBP) from 89 subjects, with ranges of 80 to 175 mmHg and 43 to 122 mmHg for SBP and DBP, respectively. In sample-wise comparison, the mean difference between the test device and the reference was 0.16 ± 5.90 mmHg (95% limits of agreement [LOA], -11.41, 11.72) in SBP and -0.07 ± 4.68 (95% LOA, -9.26, 9.10) in DBP. The test device showed a strong correlation with the reference for SBP (r = 0.94, P < 0.001) and DBP (r = 0.95, P < 0.001). There were consistent results in subject-wise comparison. CONCLUSION: The new ring-type BP measuring device showed a good correlation for SBP and DBP with minimal bias compared with an auscultatory method.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。