Development of an animal experimental model for a bileaflet mechanical heart valve prosthesis

双叶机械心脏瓣膜假体动物实验模型的建立

阅读:1

Abstract

The objective of this study was to develop a pre-clinical large animal model for the in vivo hemodynamic testing of prosthetic valves in the aortic position without the need for cardiopulmonary bypass. Ten male pigs were used. A composite valved conduit was constructed in the operating room by implanting a prosthetic valve between two separate pieces of vascular conduits, which bypassed the ascending aorta to the descending aorta. Prior to applying a side-biting clamp to the ascending aorta for proximal grafting to the aortic anastomosis, an aorta to femoral artery shunt was placed just proximally to this clamp. The heart rate, cardiac output, Vmax, transvalvular pressure gradient, effective orifice area and incremental dobutamine stress response were assessed. A dose-dependent increase with dobutamine was seen in terms of cardiac output, Vmax, and the peak transvalvular pressure gradient both in the native and in the prosthetic valve. However, the increment was much steeper in the prosthetic valve. No significant differences in cardiac output were noted between the native and the prosthetic valves. The described pre-clinical porcine model was found suitable for site-specific in-vivo hemodynamic assessment of aortic valvular prosthesis without cardiopulmonary bypass.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。