A Highly Accurate Method for Measuring Response Time of MEMS Thermopiles

一种测量MEMS热电堆响应时间的高精度方法

阅读:1

Abstract

The response time is an important parameter for thermopiles sensors, which reflects the response speed of the device. The accurate measurement of response time is extremely important to evaluate device characteristics for using them in suitable scenarios. In this work, to accurately measure the response time of thermopile sensors, an Al microheater is integrated in a MEMS thermopile as an in situ heat source. Compared with the traditional chopper measurement method for response time, this approach avoids mechanical delay induced by chopper blades. Accordingly, based on this approach, the response time of the device is measured to be 6.9 ms, while that is 12.7 ms when a chopping system is used, demonstrating that an error of at least 5.8 ms is avoided. Such an approach is quite simple to realize and provides a novel route to accurately measure the response time.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。