Solid cryogen: a cooling system for future MgB(2) MRI magnet

固态低温液体:未来MgB₂ MRI磁体的冷却系统

阅读:1

Abstract

An efficient cooling system and the superconducting magnet are essential components of magnetic resonance imaging (MRI) technology. Herein, we report a solid nitrogen (SN(2)) cooling system as a valuable cryogenic feature, which is targeted for easy usability and stable operation under unreliable power source conditions, in conjunction with a magnesium diboride (MgB(2)) superconducting magnet. The rationally designed MgB(2)/SN(2) cooling system was first considered by conducting a finite element analysis simulation, and then a demonstrator coil was empirically tested under the same conditions. In the SN(2) cooling system design, a wide temperature distribution on the SN(2) chamber was observed due to the low thermal conductivity of the stainless steel components. To overcome this temperature distribution, a copper flange was introduced to enhance the temperature uniformity of the SN(2) chamber. In the coil testing, an operating current as high as 200 A was applied at 28 K (below the critical current) without any operating or thermal issues. This work was performed to further the development of SN(2) cooled MgB(2) superconducting coils for MRI applications.