Comparison between immunoturbidimetry, size-exclusion chromatography, and LC-MS to quantify urinary albumin

免疫比浊法、尺寸排阻色谱法和液质联用法定量尿白蛋白的比较

阅读:8
作者:Aisha Shaikh, Jesse C Seegmiller, Timothy M Borland, Bradley E Burns, Paula M Ladwig, Ravinder J Singh, Rajiv Kumar, Timothy S Larson, John C Lieske

Background

The accurate and precise measurement of urinary albumin is critical, since even minor increases are diagnostically sensitive indicators of renal disease, cardiovascular events, and risk for death. To gain insights into potential measurement biases, we systematically compared urine albumin measurements performed by LC-MS, a clinically available immunoturbidimetric assay, and size-exclusion HPLC.

Conclusions

Efforts are needed to establish gold-standard traceable calibrators for clinical assays. LC-MS is a specific method to quantify albumin in native urine when concentrations exceed 20 mg/L, and therefore could be employed for standardization among assays.

Methods

We obtained unused clinical urine samples from 150 patients who were stratified by degrees of albuminuria (<20 mg/L, 20-250 mg/L, >250 mg/L) as determined by the immunoturbidimetric assay used in our clinical laboratory (Roche Hitachi 912). Urine albumin was then remeasured via LC-MS and HPLC (Accumin) assays.

Results

The immunoturbidimetric assay, calibrated using manufacturer-supplied serum-derived calibrators (Diasorin), underestimated albumin compared with LC-MS. After calibration with purified HSA, this immunoturbidimetric assay correlated well with LC-MS. HPLC overestimated albumin compared with both LC-MS and immunoturbidimetry. The current LC-MS and HPLC assays both performed poorly at concentrations <20 mg/L. Conclusions: Efforts are needed to establish gold-standard traceable calibrators for clinical assays. LC-MS is a specific method to quantify albumin in native urine when concentrations exceed 20 mg/L, and therefore could be employed for standardization among assays.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。