Inverse Pneumatic Artificial Muscles for Application in Low-Cost Ventilators

用于低成本呼吸机的反向气动人工肌肉

阅读:2

Abstract

The procurement and maintenance cost of high-end ventilators preclude their stockpiles sufficient for the mass emergency situations. Therefore, there is a significant demand for mechanical ventilators in such situations. Herein, a low-cost, portable, yet high-performance design for a volume-controlled mechanical ventilator is proposed. Pneumatic artificial muscles, such as air cylinders, are used in the inverse mode of operation to achieve mechanical ventilation. With the current design, the two fundamental modes of operation (controlled mode and assisted mode) are demonstrated. Unlike most intensive care unit ventilators, the proposed device does not need a high-pressure air pipeline to operate. The device is capable of mechanical ventilation for respiration rate ranging from 10 to 30 b min(-1) with a tidal volume (VT) range of 150-1000 mL and the I:E ratio of 1:1-1:5. A total cost of less than $400 USD is achieved to make one device. The cost to produce the device in larger volumes can be estimated to be less than $250 USD.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。