Insertion of peripherally inserted central catheters with intracavitary electrocardiogram guidance: A randomized multicenter study in China

经外周静脉置入中心静脉导管并进行腔内心电图引导:一项中国多中心随机研究

阅读:1

Abstract

INTRODUCTION: Ultrasound-guided venipuncture and tip location by intracavitary electrocardiogram have many advantages during the insertion of peripherally inserted central catheters, both in terms of safety and cost-effectiveness. Recently, a new tip-conductive peripherally inserted central catheters and new Doppler ultrasound device integrated with intracavitary electrocardiogram have been introduced into clinical practice in China. A randomized multicenter study (clinical trial no. NCT03230357) was performed to verify the feasibility and accuracy of intracavitary electrocardiogram, as performed with this new peripherally inserted central catheters and device. METHODS: Our study enrolled a total of 2250 adult patients in 10 different Chinese hospitals. The patients were randomly assigned to either the study group (intracavitary electrocardiogram) or the control group (anatomical landmark guidance) in a 2:1 allocation. Ultrasound was used in both groups for venipuncture and tip navigation. All patients underwent chest X-ray after the procedure to verify the position of the catheter tip. RESULTS: No insertion-related complications were reported in either group. In the study group, first-attempt successful tip location was 91.7% (95% confidence interval: 90.3%-93.1%), significantly higher than 78.9% (95% confidence interval: 76.0%-81.9%) observed in the control group (p < 0.001). As evaluated by post-procedural chest X-ray, tip location in the study group had a sensitivity of 99.3% (95% confidence interval: 98.8%-99.7%), significantly higher than 86.8% (95% confidence interval: 84.4%-89.2%) observed in the anatomical landmark group (p < 0.001). CONCLUSION: These results indicated that during peripherally inserted central catheters insertion in adult patients, tip location with intracavitary electrocardiogram guidance, as carried out by a new tip-conductive peripherally inserted central catheters and intracavitary electrocardiogram integrated ultrasound device, was more effective and more accurate than tip location using anatomical landmarks.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。