Left Atrial Appendage Management with the Watchman Device during Hybrid Ablation of Atrial Fibrillation

在房颤混合消融术中使用Watchman装置进行左心耳管理

阅读:1

Abstract

BACKGROUND: In the recent ESC/EACTS guidelines, left atrial appendage (LAA) occlusion or exclusion in patients undergoing (thoracoscopic) atrial fibrillation (AF) ablation surgery is recommended. The Watchman device (WD, Boston Scientific, Minnesota) has proved to reduce the risk of thromboembolic events by closing of the LAA, yet no data exist on WD implantation during surgical AF ablation. The objective is to determine if WD implantation is safe and feasible in a hybrid AF ablation setting (i.e., combination of thoracoscopic epicardial surgical and endocardial catheter ablation) and could become subject of further testing to serve as a bail-out in cases in which surgical LAA occlusion methods cannot be applied, due to, for example, severe adhesions. METHODS: In this prospective, single center, pilot study, 10 consecutive patients undergoing a hybrid ablation qualifying for LAA exclusion (CHA(2)DS(2)-VASc ≥ 1) were included. At the end of the hybrid ablation, the LAA was occluded endocardially using the WD. The feasibility endpoint was successful implantation. The safety endpoint concerned major complications. RESULTS: One patient was excluded and replaced because the LAA was insufficiently visible on transesophageal echocardiography. In 10/11 patients, device delivery was successful (mean time: 35 minutes). No major complications occurred. Transesophageal echocardiography after 6 weeks and 6 months showed successful occlusion of the LAA without significant peridevice flow. CONCLUSION: Implantation of the WD seems to be feasible and safe in the setting of hybrid AF ablation and could be an alternative to epicardial occlusion in surgical AF ablation procedures. Larger studies are required to confirm these findings. This trial is registered with NCT02471131.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。