Calibration of Time-Interleaved Errors in Digital Real-Time Oscilloscopes

数字实时示波器中时间交错误差的校准

阅读:1

Abstract

A channel mismatch calibration method is proposed for use in time-interleaved digital real-time oscilloscope (DRTO) applications. Linear equations are derived using Fourier transforms of the separated signals from each of the time-interleaved analog-to-digital converters (TIADCs). Thus the errors in the TIADCs can be easily calibrated by inversion of a matrix, as opposed to most previous work where additional filters are employed. The calibration accuracy of the proposed method is limited only by the noise produced after the TIADC circuitry, while other methods depend on the filter design. A transfer function measurement method is then proposed for application to commercially available DRTOs. Two-tone signals are measured using DRTOs from various suppliers to validate the proposed method. The occurrence of signals at spurious frequencies is considerably reduced, as demonstrated by the calibrated results.