Study of Power Equivalent Continuous Approximation Based on the Recent Consensus Recommendations for Brain Tumor Imaging with Pulsed Chemical Exchange Saturation Transfer at 3T

基于最新共识建议,研究3T脉冲化学交换饱和转移脑肿瘤成像中功率等效连续近似方法。

阅读:1

Abstract

The quantitative analysis of pulsed-chemical exchange saturation transfer (CEST) using a full model-based method is computationally challenging, as it involves dealing with varying RF values in pulsed saturation. A power equivalent continuous approximation of B(1) power was usually applied to accelerate the analysis. In line with recent consensus recommendations from the CEST community for pulsed-CEST at 3T, particularly recommending a high RF saturation power (B(1) = 2.0 µT) for the clinical application in brain tumors, this technical note investigated the feasibility of using average power (AP) as the continuous approximation. The simulated results revealed excellent performance of the AP continuous approximation in low saturation power scenarios, but discrepancies were observed in the z-spectra for the high saturation power cases. Cautions should be taken, or it may lead to inaccurate fitted parameters, and the difference can be more than 10% in the high saturation power cases.