Nature of the Intracellular-contrast-enhancing Fat-saturated T1-weighted Gradient-echo (ICE-TIGRE) Sequence: A Fat-suppressed T1-weighted Technique with Motion-sensitised Driven-equilibrium for Improved Contrast Enhancement in Liver Imaging

细胞内对比增强脂肪抑制T1加权梯度回波(ICE-TIGRE)序列的特性:一种采用运动敏感驱动平衡的脂肪抑制T1加权技术,可提高肝脏成像的对比度增强效果

阅读:2

Abstract

Gadoxetic acid is both an extracellular- and hepatocyte-specific contrast agent. Signals from the extracellular space may lower the contrast between lesions and the surrounding hepatic parenchyma. To improve hepatocyte-specific enhancement, we developed an intracellular contrast-enhancing fat-saturated T1-weighted gradient-echo nature of the sequence (ICE-TIGRE). It incorporates the motion-sensitized driven-equilibrium (MSDE) pulse to suppress signals from the blood flow. We investigated the optimal ICE-TIGRE scanning parameters, i.e., the order of the MSDE and the fat saturation pulses, the duration time, and the b value of the MSDE pulse, using a phantom and three volunteers without applying gadoxetic acid. ICE-TIGRE successfully increased the contrast between the liver parenchyma and the portal vein. To maintain fat saturation, the preparation pulse order should be MSDE-fat saturation. A duration time of 21 ms should be applied to minimize the effect of the T2 factor on the T1 contrast, and a b value of 60 s/mm(2) should be applied to maximize the diffusion contrast for ICE-TIGRE with the imaging system used in this study.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。