Force measurement of insertion of cochlear implant electrode arrays in vitro: comparison of surgeon to automated insertion tool

体外人工耳蜗电极阵列插入力测量:外科医生与自动插入工具的比较

阅读:1

Abstract

CONCLUSIONS: We have demonstrated that an automated insertion tool (i.e. a robot) can be used to duplicate a complex surgical motion in inserting cochlear implant (CI) electrode arrays via the 'advance-off-stylet' (AOS) technique. As compared with human operators, the forces generated by the robot were slightly larger but the robot was more reliable (i.e. less force maxima). OBJECTIVES: We present force data collected during CI electrode insertion by human operators and by an automated insertion tool. MATERIALS AND METHODS: Using a three-dimensional, anatomically correct, translucent model of the scala tympani chamber of the cochlea, CI electrodes were inserted either by one of three surgeons (26 insertions) or by the robotic insertion tool (8 insertions). Force was recorded using a load beam cell calibrated for expected forces of <0.1 Newtons (N). The insertions were also videotaped to allow correlation of force with depth of penetration into the cochlea and speed of insertion. RESULTS: Average insertion force used by the surgeons was 0.004+/-0.001 N and for the insertion tool it was 0.005+/-0.014 N (p<0.00001, Student's t test). While the average insertion force of the automated tool was larger than that of the surgeons, the surgeons did have intermittent peaks during the AOS component of the insertion (between 120 degrees and 200 degrees ).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。