High-Throughput Testing of Urogenital and Extragenital Specimens for Detection of Chlamydia Trachomatis and Neisseria Gonorrhoeae with Cobas(®) CT/NG

利用 Cobas® CT/NG 对泌尿生殖道和生殖道外标本进行高通量检测,以检测沙眼衣原体和淋病奈瑟菌

阅读:1

Abstract

We compared the analytical and clinical performance of cobas(®) CT/NG for use on the Cobas(®) 6800/8800 Systems with the Cobas(®) 4800 CT/NG Test from urogenital and extragenital specimens in over 12,000 specimens from both male and female subjects in Germany and the United States. The analytical sensitivity was ≤40 EB/ml for Chlamydia trachomatis (CT) and ≤1 CFU/ml for Neisseria gonorrhoeae (NG). Using clinical specimens, the overall percent agreement with the Cobas(®) 4800 CT/NG Test was >98.5%. Across urogenital specimens, there were 93 discrepant specimens; 76 (93.8%) of 81 CT discrepant specimens were 6800+/4800- and 10 (83.3%) of 12 NG discrepant specimens were 6800+/4800-. Sequencing verified CT results for 45 (61.6%) of 73 samples positive by 6800 and 1 (20%) of 5 positive by 4800. Similarly, 7 (70.0%) of 10 NG samples positive by 6800 and 1 of 2 positive by 4800 were confirmed by sequencing. Among discrepant extragenital specimens (all 6800+/4800-), 7 (50%) of 14 oropharyngeal and 23 (76.7%) of 30 anorectal CT discordant samples were confirmed as CT positive by sequencing; all 8 anorectal and 20 (90.9%) of 22 oropharyngeal NG discordant results were also confirmed as NG positive. In conclusion, Cobas(®) CT/NG for use on the Cobas(®) 6800/8800 Systems provides high-throughput automated solutions for sexually transmitted infection (STI) screening programs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。