Effect of different kinematics and perforation diameter on integrated electronic apex locator accuracy in detecting root canal perforations

不同运动学特征和穿孔直径对集成电子根尖定位仪检测根管穿孔精度的影响

阅读:1

Abstract

This study aimed to evaluate the precision and reliability of the integrated electronic apex locator in determining various root perforation diameters and to examine the impact of instrumentation kinematics on perforation site detection. One hundred extracted single-root mandibular premolars were artificially perforated at five distinct diameters (0.25-1.25 mm), located 5 mm above the root apex. The teeth were then divided into 10 groups based on perforation size and kinematic subgroup (rotary, reciprocation; n = 10/subgroup). Actual distances to the perforation site (actual length) were measured under a dental operating microscope. Electronic distances (electronic length) were determined using an endodontic motor with an integrated electronic apex locator using rotary or reciprocating instrumentation kinematics during root canal treatment. To quantify the actual length to the electronic length discrepancies, individual tooth measurements were compared. The integrated electronic apex locator failed to identify perforations ≤ 0.50 mm in diameter. Electronic length measurements were similarly accurate for 1.00 and 1.25 mm perforations but more precise than for 0.75 mm perforations. Instrumentation kinematics did not affect the electronic length measurement accuracy. Integrated electronic apex locators failed to detect small perforations during root canal treatment. The perforation diameter influenced the detection accuracy of integrated electronic apex locators, whereas kinematics had no effect.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。