The contribution of interventional cardiology procedures to the population radiation dose in a 'health-care level I' representative region

介入心脏病手术对“一级医疗机构”代表性区域人群辐射剂量的贡献

阅读:1

Abstract

This study evaluates per-procedure, collective and per capita effective dose to the population by interventional cardiology (IC) procedures performed during 2002-11 at the main hospital of Aosta Valley Region that can be considered as representative of the health-care level I countries, as defined by the UNSCEAR, based on its socio-demographic characteristics. IC procedures investigated were often multiple procedures in patients older than 60 y. The median extreme dose-area product values of 300 and 22 908 cGycm(2) were found for standard pacemaker implantation and coronary angioplasty, respectively, while the relative mean per-procedure effective dose ranged from 0.7 to 47 mSv. A 3-fold increase in frequency has been observed together with a correlated increase in the delivered per capita dose (0.05-0.27 mSv y(-1)) and the collective dose (5.8-35 man Sv y(-1)). Doses increased particularly from 2008 onwards mainly because of the introduction of coronary angioplasty procedures in the authors' institution. IC practice contributed remarkably in terms of effective dose to the population, delivering ∼10% of the total dose by medical ionising radiation examination categories.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。