Determination of 234U/238U, 235U/238U and 236U/238U isotope ratios in urine using sector field inductively coupled plasma mass spectrometry

利用扇形场电感耦合等离子体质谱法测定尿液中 234U/238U、235U/238U 和 236U/238U 同位素比值

阅读:1

Abstract

Quantification of the isotopic composition of uranium in urine at low levels of concentration is important for assessing both military and civilian populations' exposures to uranium. However, until now there has been no convenient, precise method established for rapid determination of multiple uranium isotope ratios. Here, the authors report a new method to measure (234)U/(238)U, (235)U/(238)U and (236)U/(238)U. It uses solid-phase chelation extraction (via TRU columns) of actinides from the urine matrix, followed by measurement using a magnetic sector field inductively coupled plasma mass spectrometer (SF-ICP-MS-Thermo Element XR) equipped with a high-efficiency nebulizer (Apex PFA microflow) and coupled with a membrane desolvating nebulizer system (Aridus II™). This method provides rapid and reliable results and has been used successfully to analyse Certified Reference Materials.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。