Pediatric eye movement-based perimetry: progress, pitfalls, and prospects

儿童眼动视野检查:进展、不足和前景

阅读:1

Abstract

INTRODUCTION: Eye movement-based perimetry (EMP) is a promising, non-invasive approach for visual field assessment, particularly in pediatric populations where standard automated perimetry often fails. However, completion rates in prior pediatric EMP studies have ranged from 41 to 81%, and reasons for unsuccessful testing are seldom reported. OBJECTIVE: In this perspective article, we aim to highlight practical barriers and design gaps in EMP systems for children, with a focus on clinical use. OBSERVATIONS: From our clinical experience with testing two commercially available EMP systems in children (21 patients with brain tumors and 19 age-matched controls), we observed recurring challenges, including poor ergonomic fit, inadequate calibration of eye tracker, and insufficient attention management strategies. These issues frequently led to data loss and incomplete tests, underscoring the gap between current technology and pediatric clinical needs. We outline solutions informed by technological development, vision science and clinical ophthalmology. CONCLUSION: Pediatric testing experience must inform EMP design to ensure accessibility and reliability. Our observations highlight the need for clinician-engineer-scientist collaboration, with innovations likely to benefit not only children but also adults with similar testing challenges.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。